Medizinische Fachkräfte

Actions

ProMRI AFFIRM

Actions

Study Design

Key Result 1

SADE free rate: 100% free of MRI and pacing system related serious adverse device effects 1

Key result 2:

Atrial pacing threshold: 99% of patients experienced < 0.5 V atrial pacing threshold increase post-MRI11

Key Result 3:

Ventricular pacing threshold: 100% of patients experienced < 0.5 V ventricular pacing threshold increase post-MRI1 P-wave sensing attenuation: 100% of patients experienced P-waves above 1.5 mV — 99.4% of patients experienced < 50% P-wave sensing attenuation R-wave sensing attenuation: 100% of patients experienced R-waves above 5 mV — 99.5% of patients experienced < 50% R-wave sensing attenuation

Clinical Relevance