Medizinische Fachkräfte

Actions

ProMRI (Phase B)

Actions

Study Design

Key result 1

SADE free rate: 99.6% free of MRI and pacing system related serious adverse device effects1

Key Result 2

Atrial pacing threshold: 100% of patients experienced < 0.5 V atrial pacing threshold increase post-MRI1

Key Result 3

Ventricular pacing threshold: 100% of patients experienced < 0.5 V ventricular pacing threshold increase post-MRI1 P-wave sensing attenuation: 98.2% of patients experienced P-waves above 1.5 mV — 99.4% of patients experienced < 50% P-wave sensing attenuation R-wave sensing attenuation: 100% of patients experienced R-waves above 5 mV — 100% of patients experienced < 50% R-wave sensing attenuation

Clinical Relevance

Clinical safety of the ProMRI pacemaker system in patients subjected to head and lower lumbar 1.5-T magnetic resonance imaging scanning conditions