Effectiveness and Cost Of ICD Follow-up Schedule With Telecardiology

Study Design

  • Randomized, interventional, multicenter study
  • To compare the safety of continuous Home Monitoring versus conventional ambulatory follow-up of ICDs.
  • 433 patients at 43 centers in France

Key Result 1

Remote follow-up with BIOTRONIK Home Monitoring® is as safe as conventional ambulatory follow-up.

圖片
Proportion of Patients

Key Result 2

Continuous monitoring with BIOTRONIK Home Monitoring® reduced the rate of patients who experienced inappropriate shocks significantly by 52%. 72% fewer hospitalizations due to inappropriate shocks.

圖片
Inappropriate shocks

Key Result 3

BIOTRONIK Home Monitoring® reduced the number of charged shocks by 76%. Reducing shocks saved 7.9 months of battery longevity.

圖片
Number of Capacitors Charges

Clinical relevance

  • The ECOST study shows that BIOTRONIK Home Monitoring of ICDs as a safe alternative to in-office follow-ups. It positively impact on patient outcome , specially when it comes to reducing the risk of inappropriate shocks – a very unpleasant experience.
  • The value of BIOTRONIK Home Monitoring for continuous remote monitoring, through its daily transmissions, is essential for clinics to better manage their patients health condition.
Reference no.
  • NCT00989417
1° Endpoints
  • The proportion of patients who experienced ≥ 1 MAE, including death from any cause, cardiovascular and procedure or device related MAE.
2° Endpoints
  • Appropriate and inappropriate shock deliveries
  • ICD capacitor charges
  • Impact on battery longevity
Clinical Sites
  • 43 centers
Sample Size
  • 433 patients
Inclusion Criteria
  • Indication for single or dual chamber ICD
Main Exclusion Criteria #1
  • NYHA class IV
  • Pregnant woman or woman who plan to become pregnant during the trial
  • Patient whose medical situation is not stable
  • Presence of any disease, other than patient's cardiac disease, associated with reduced likelihood of survival for the duration of the trial, e.g. cancer, uraemia (urea > 70mg/dl or creatinine >3mg/dl), liver failure, etc.
  • Age
  • Patient unable to handle Home Monitoring system correctly
  • The patient is not willing and able to comply with the protocol
  • Change of residence expected during study
Main Exclusion Criteria #2
  • Insufficient GSM coverage at patient's home
  • Participation in another clinical study
  • Patient unwilling to sign the consent for participation.
Devices
  • ICDs equipped with BIOTRONIK Home Monitoring
Study FlowchartStudy Flowchart
Follow-Up
  • 27 months follow-up
Study Duration
  • January 2007 – October 2010 (Final data collection date for primary outcome measure)
Principal Investigators
  • Salem S. Kacet, Prof. Dr. , CHRU de Lille – Hospital Cardiologique, France

Download Section