Orsiro Mission

Vascular Intervention // Coronary
Drug-Eluting Stent System

 
Orsiro Mission
 
The next level of deliverability for the outstanding Orsiro stent.
Learn more at orsiro-mission.com.
 
 


The next level of deliverability
 
Ultrathin struts
 
Outstanding patient outcomes

 


 

 

 

Orsiro Mission

Indicated for discrete de novo stenotic lesions and in-stent restenotic lesions1.

Technical Data

Stent
Stent material Cobalt chromium, L-605
Strut thickness ø 2.25 – 3.0 mm: 60 μm (0.0024”);
ø 3.50 – 4.0 mm: 80 μm (0.0031”)
Passive coating proBIO® (Amorphous Silicon Carbide)
Active coating BIOluteTM bioabsorbable Poly-L-Lactide (PLLA) eluting a limus drug
Drug dose 1.4 μg/mm2
Delivery System
Catheter type Rapid exchange
Recommended guide catheter 5F (min. I.D. 0.056")
Guide wire diameter 0.014"
Usable catheter length 140 cm
Balloon material Semi crystalline polymer material
Coating (Distal shaft) Hydrophilic
Coating (Proximal shaft) Hydrophobic
Marker bands Two swaged platinum-iridium markers
Lesion entry profile 0.017"
Distal shaft diameter 2.7F: ø 2.25 – 3.0 mm; 2.9F: ø 3.5 - 4.0 mm
Proximal shaft diameter 2.0F
Nominal pressure (NP) 10 atm
Rated burst pressure (RBP) 16 atm
Use Before Date (UBD) 24 months
Temperature Between 15°C (59°F) and 25°C (77°F), short term excursions between 10°C (50°F) and 40°C (104°F) are allowed
Specifications
Double Helix Stent Design
Nominal diameter (mm)    2.25 2.5 2.75 3.0 3.5 4.0
Strut thickness (μm) 60 60 60 60 80 80
Strut width (μm) 75 75 75 75 85 85
Amount of connectors 3 3 3 3 3 3
Amount of crowns at end 8 8 8 8 8 8
Maximal Expansion and Stent Strut Opening
Nominal diameter (mm) 2.25 2.5 2.75 3.0 3.5 4.0
Maximal expansion diameter (mm) 3.5 3.5 3.5 3.5 4.5 4.5
Maximal diameter of expanded stent cell (mm) 3.59 3.59 3.59 3.59 4.42 4.42
Stent strut opening diameter at NP* (mm) 0.79 0.92 0.92 0.92 4.42 4.42
Nominal outer diameter of the stent at NP (mm) 2.37 2.62 2.87 3.12 3.66 4.16
*Mean of the largest possible opening diameter within a stent cell at NP
Ordering Information
  9 13 15 18  22 26 30 35 40
2.25 419101 419107 419113 419119 419125 419131 419137 419143 419149
2.5 419102 419108 419114 419120 419126 419132 419138 419144 419150
2.75 419103 419109 419115 419121 419127 419133 419139 419145 419151
3.0 419104 419110 419116 419122 419128 419134 419140 419146 419152
3.5 419105 419111 419117 419123 419129 419135 419141 419147 419153
4.0 419106 419112 419118 419124 419130 419136 419142 419148 419154

Contact

contact
Country/Region

1 As per IFU