Scoring Balloon Catheter
Blimp

High CTO penetration capability with ultra low profile balloon

Enhanced dilatation capability due to scoring element

Hydrophilic coating for advanced crossability

Blimp
CTO Scoring Balloon Catheter
Indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion.*
Blimp is designed to cross severe lesions due to its ultra low profile balloon and internal stiffening wire reaching from the hub to the tip, resulting in high penetration capability. It is a unique PTCA device, designed to modify calcium within the lesion as a result of its very small, yet very high rated burst scoring balloon combined with the focused scoring effect of the guide wire.

Technical Data
CTO Scoring Balloon Catheter | |
---|---|
Nominal Pressure (NP) | 25 atm |
Rated Burst Pressure (RBP) | 30 atm |
Guide wire compatibility | 0,014" |
Balloon length | 5 mm |
Inflated balloon diameter | 0,6 mm |
Shaft diameter | 1,5 F |
Rx section length | 4 mm |
Balloon radiopaque marker | Center of balloon |
Distal coating | Hydrophilic, 25 cm |
Exit markers | 95 cm and 105 cm |
Ordering Information | Ordering article number | IMDS article number | Guiding catheter compatibility | Usable length | ||||||||||
451670 | BRX0605 | 5 F | 150 cm |
Contact
*Indication as per IFU