Technical Data
Stent | |
---|---|
Stent material | Cobalt chromium, L-605 |
Strut thickness | ø 2.25 – 3.0 mm: 60 μm (0.0024”); ø 3.50 – 4.0 mm: 80 μm (0.0031”) |
Passive coating | proBIO® (Amorphous Silicon Carbide) |
Active coating | BIOluteTM bioabsorbable Poly-L-Lactide (PLLA) eluting a limus drug |
Drug dose | 1.4 μg/mm2 |
Delivery System | |
---|---|
Catheter type | Rapid exchange |
Recommended guide catheter | 5F (min. I.D. 0.056") |
Guide wire diameter | 0.014" |
Usable catheter length | 140 cm |
Balloon material | Semi crystalline polymer material |
Coating (Distal shaft) | Hydrophilic |
Coating (Proximal shaft) | Hydrophobic |
Marker bands | Two swaged platinum-iridium markers |
Lesion entry profile | 0.017" |
Distal shaft diameter | 2.7F: ø 2.25 – 3.0 mm; 2.9F: ø 3.5 - 4.0 mm |
Proximal shaft diameter | 2.0F |
Nominal pressure (NP) | 10 atm |
Rated burst pressure (RBP) | 16 atm |
Use Before Date (UBD) | 24 months |
Temperature | Between 15°C (59°F) and 25°C (77°F), short term excursions between 10°C (50°F) and 40°C (104°F) are allowed |
Specifications | |||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Double Helix Stent Design | |||||||||||||||||
Nominal diameter (mm) | 2.25 | 2.5 | 2.75 | 3.0 | 3.5 | 4.0 | |||||||||||
Strut thickness (μm) | 60 | 60 | 60 | 60 | 80 | 80 | |||||||||||
Strut width (μm) | 75 | 75 | 75 | 75 | 85 | 85 | |||||||||||
Amount of connectors | 3 | 3 | 3 | 3 | 3 | 3 | |||||||||||
Amount of crowns at end | 8 | 8 | 8 | 8 | 8 | 8 | |||||||||||
Maximal Expansion and Stent Strut Opening | |||||||||||||||||
Nominal diameter (mm) | 2.25 | 2.5 | 2.75 | 3.0 | 3.5 | 4.0 | |||||||||||
Maximal expansion diameter (mm) | 3.5 | 3.5 | 3.5 | 3.5 | 4.5 | 4.5 | |||||||||||
Maximal diameter of expanded stent cell (mm) | 3.59 | 3.59 | 3.59 | 3.59 | 4.42 | 4.42 | |||||||||||
Stent strut opening diameter at NP* (mm) | 0.79 | 0.92 | 0.92 | 0.92 | 4.42 | 4.42 | |||||||||||
Nominal outer diameter of the stent at NP (mm) | 2.37 | 2.62 | 2.87 | 3.12 | 3.66 | 4.16 | |||||||||||
*Mean of the largest possible opening diameter within a stent cell at NP |
Ordering Information | |||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|
9 | 13 | 15 | 18 | 22 | 26 | 30 | 35 | 40 | |||
2.25 | 419101 | 419107 | 419113 | 419119 | 419125 | 419131 | 419137 | 419143 | 419149 | ||
2.5 | 419102 | 419108 | 419114 | 419120 | 419126 | 419132 | 419138 | 419144 | 419150 | ||
2.75 | 419103 | 419109 | 419115 | 419121 | 419127 | 419133 | 419139 | 419145 | 419151 | ||
3.0 | 419104 | 419110 | 419116 | 419122 | 419128 | 419134 | 419140 | 419146 | 419152 | ||
3.5 | 419105 | 419111 | 419117 | 419123 | 419129 | 419135 | 419141 | 419147 | 419153 | ||
4.0 | 419106 | 419112 | 419118 | 419124 | 419130 | 419136 | 419142 | 419148 | 419154 |