Product Details

Highly Pushable Coaxial Shaft Design
The low profile and 3.8 F coaxial catheter shaft design facilitates access to most distal lesions and minimizes access site complications.1
Tip
Smooth tapered tip entry profile promoting enhanced lesion crossing.
Controlled Compliance
The innovative balloon design and material enable a low average compliance rate. This predictable radial balloon expansion suggests to minimize the risk of dissection.

Patchwork Coating for Enhanced Crossability
The penta-fold balloon is fully coated when folded and only partly coated when inflated. This enables enhanced crossability while minimizing slippage.
Technical Data
Passeo-18 Balloon Catheter | |
---|---|
Catheter type | OTW |
Recommended guide wire | 0.018" |
Tip | Short, tapered |
Balloon material | SCP (semi-crystalline polymer), controlled compliance (4 - 8 %) |
Balloon folding | 5-fold |
Balloon coating | Hydrophobic patchwork coating |
Balloon markers | 2 swaged markers (zero profile) |
Sizes | ø 2.0 - 7.0 mm; L: 20 - 200 mm |
Shaft | 3.8 F, 3.9 F (ø 6.0/7.0 mm x 170 - 200 mm); coaxial design |
Usable length | 90, 130 and 150 cm |
Compliance Chart
Balloon Diameter x Length (mm) | |||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Nominal Pressure | atm3 | 6 | 6 | 6 | 6 | 6 | 6 | 6 | 6 | 6 | 6 | 6 | 6 | 6 | 6 | 6 | |||
(NP) | ø (mm) | 2.0 | 2.0 | 2.5 | 2.5 | 3.0 | 3.0 | 3.5 | 3.5 | 4.0 | 4.0 | 5.0 | 5.0 | 5.0 | 6.0 | 7.0 | |||
Rated Burst Pressure | atm3 | 15 | 14 | 15 | 14 | 15 | 14 | 15 | 14 | 15 | 13 | 15 | 12 | 13 | 12 | 12 | |||
(RBP) | ø (mm) | 2.1 | 2.1 | 2.6 | 2.6 | 3.2 | 3.2 | 3.7 | 3.7 | 4.3 | 4.2 | 5.3 | 5.2 | 5.2 | 6.2 | 7.3 |
Ordering Information
Contact
1 Bosiers M, Deloose K, Callaert J, et al. 4-French-compatible endovascular material is safe and effective in the treatment of femoropopliteal occlusive disease: results of the 4-EVER trial. J Endovasc Ther. 2013; 20: 746-756.
2 8 weeks pre-order only
3 1 atm = 1.013 bar