Passeo-18

Image
Image of Passeo-18
Passeo-18 PTA Balloon
Image
Image of Passeo-18
Passeo-18 PTA Balloon
Image
Image of Passeo-18
Passeo-18 PTA Balloon
Image
Image of Passeo-18
Passeo-18 PTA Balloon

Passeo®-18

PTA BALLOON

Indicated for lower limb arteries. Treatment of arterio-venous dialysis fistulae. The Passeo®-18 features a coaxial shaft design with a strong inner shaft and a flexible outer shaft creating a highly pushable and deliverable system. Its wide array of available sizes enables treatment of both inflow and outflow vessels.


KEY FACTS
  • Highly pushable coaxial shaft design for most distal lesion access
  • Controlled compliant balloon for predictable expansion
  • Low profile, wide range: ø 2 - 7 mm and 20 - 200 mm balloon length
  • Patchwork coating for enhanced crossability

 

Product Details

Passeo-18 image

Highly Pushable Coaxial Shaft Design

The low profile and 3.8 F coaxial catheter shaft design facilitates access to most distal lesions and minimizes access site complications.1


Tip

Smooth tapered tip entry profile promoting enhanced lesion crossing.


Controlled Compliance

The innovative balloon design and material enable a low average compliance rate. This predictable radial balloon expansion suggests to minimize the risk of dissection.


Passeo-18 image

Patchwork Coating for Enhanced Crossability

The penta-fold balloon is fully coated when folded and only partly coated when inflated. This enables enhanced crossability while minimizing slippage.


Technical Data

Passeo®-18 Balloon Catheter
Catheter type
OTW
Recommended guide wire
0.018"
Tip
Short, tapered
Balloon material
SCP (semi-crystalline polymer), controlled compliance (4 - 8 %)
Balloon folding
5-fold
Balloon coating
Hydrophobic patchwork coating
Balloon markers
2 swaged markers (zero profile)
Sizes
ø 2.0 - 7.0 mm; L: 20 - 200 mm
Shaft
3.8 F, 3.9 F (ø 6.0/7.0 mm x 170 - 200 mm); coaxial design
Usable length
90, 130 and 150 cm

Compliance Chart

Balloon Diameter x Length (mm)
Nominal Pressure atm3 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
(NP) ø (mm) 2.0 2.0 2.5 2.5 3.0 3.0 3.5 3.5 4.0 4.0 5.0 5.0 5.0 6.0 7.0
Rated Burst Pressure atm3 15 14 15 14 15 14 15 14 15 13 15 12 13 12 12
(RBP) ø (mm) 2.1 2.1 2.6 2.6 3.2 3.2 3.7 3.7 4.3 4.2 5.3 5.2 5.2 6.2 7.3

Ordering Information


Contact

contact
Country/Region

1 Bosiers M, Deloose K, Callaert J, et al. 4-French-compatible endovascular material is safe and effective in the treatment of femoropopliteal occlusive disease: results of the 4-EVER trial. J Endovasc Ther. 2013; 20: 746-756.

2 8 weeks pre-order only

3 1 atm = 1.013 bar

Passeo is a trademark or registered trademark of the BIOTRONIK Group of Companies.