Skip to main content
  • Image

Pantera® Pro

High push. High cross coronary balloon.

Indicated for dilatation of coronary artery or bypass graft stenosis.a

Category

Vascular Intervention Coronary CVI-Balloon Catheters

Product Highlights

Image

Better Crossability²

Slim shoulders coupled with hydrophilic coatingb facilitate better crossing2

Image

43% less friction³ during simultaneous use

Reduced distal shaft diameter for improved 6F simultaneous use compatibilityc1

Image

38% more pushability⁴

Enhanced Force Transmission (EFT) shaft for unmatched pushability1 

Technical Data

Proximal shaft  
Design Hypotube design
Diameter 2.0F
Shaft markers 92 cm and 102 cm from tip
Distal shaft  
Guiding catheter 5F (min. I.D. 0.056” / 1.42 mm)
Guide wire diameter 0.014”
Lesion entry profile 0.017”
Usable length 140 cm
Balloon material Semi Crystalline Co-Polymer
Balloon folding ø 1.25 - 1.5 mm: Two-fold; ø 2.0 - 4.0 mm: Tri-fold
Balloon markers Platinum-Iridium: ø 1.25 - 1.5 mm one marker; ø 2.0 - 4.0 mm  two markers
Coating distal shaft Hydrophilic (end of balloon to Guide Wire (GW) exit port)
Balloon and tip coating ø 1.25 - 2.0 mm: Hydrophilic
ø 2.50 - 4.0 mm: Hydrophobic
Simultaneous use compatibility 6F guiding catheter (min. I.D. 0.070” / 1.78 mm), up to ø 3.5 mm
Diameter 2.6F (ø 1.25 - 2.0 mm); 2.7F (ø 2.5 - 3.5 mm); 2.9F (ø 4.0 mm)

Compliance Chart

    Balloon diameter x length (mm)
    ø 1.25 x
6-20
ø 1.50 x
6-20
ø 2.00 x
10-30 
ø 2.50 x
10-30
 ø 3.00 x
10-30
 ø 3.50 x
10-30 
 ø 4.00 x
10-30 
Nominal Pressure
(NP)
atm* 7 7 7 7 7 7 7
ø (mm) 1.24 1.49 2.01 2.49 3.08 3.62 3.95
Rated Burst
Pressure (RBP)
atm* 14 14 14 14 14 14 14
ø (mm) 1.37 1.72 2.23 2.93 3.50 4.06 4.55
*1 atm = 1.013 bar

Ordering Information

  Balloon
ø (mm)
Catheter length 140 cm
Balloon length (mm)
     
    6 10 15 20 25 30
5F 1.25 393289 393291 393298 393305 - -
1.50 393290 393292 393299 393306 - -
2.00 - 393293 393300 393307 393312 393317
2.50 - 393294 393301 393308 393313 393318
3.00 - 393295 393302 393309 393314 393319
3.50 - 393296 393303 393310 393315 393320
4.00 - 393297 393304 393311 393316 393321

How can we help you?

I prefer to be contacted by
*Fields are mandatory.

References

a. Indication as per IFU; b. 1.25-2.0 mm diameter; c. Any combination of two diameters not larger than 3.5 mm within a 6F guiding catheter with a minimal inner diameter of 0.070”/1.78 mm.

1. BIOTRONIK data on file; 2. Bench test with 1.5/15mm size when compared to Mini Trek and Emerge, BIOTRONIK data on file; 3. Bench test with 3.0/20mm size when compared to Trek, BIOTRONIK data on file; 4. Bench test with 1.5/15mm size when compared to Mini Trek, BIOTRONIK data on file.

Pantera is a trademark or registered trademark of the BIOTRONIK Group of Companies.